Registry · equipment · Various (e.g. Esco MIRI, Cook MINC, K-Systems G-series, Planer BT37)
Benchtop incubators (dry / humidified, multi-chamber)
Small-chamber incubators with rapid recovery, often dry with premixed gas, one or two dishes per chamber.
What it is and how it's used
Fast recovery after opening and per-chamber isolation make them forgiving of door openings, but small volumes and dry operation make drop osmolality and gas supply purity critical.
Typical failure modes
- Displayed vs actual CO₂/temperature drift; a chamber lid seal failing.
- Gas supply issue: cylinder change, regulator, in-line filter saturation → pH shift across all chambers.
- Dry chamber + small drops → osmolality rise over long culture.
- Post-service set-point error.
What to check first
- Independent CO₂ (and O₂) measurement per chamber; dish-level temperature by calibrated probe.
- Gas cylinder change dates and filter change dates vs KPI change.
- Osmolality of medium after culture in that chamber.
- Compare KPIs by chamber/incubator.
IFU and practice notes
Follow the manufacturer's IFU for calibration intervals, humidification, gas requirements and cleaning agents. Validate before first patient use and after service.
Neutral, and the IFU wins
This page summarises common experience across labs and is not manufacturer guidance. Always follow the current instructions for use for the exact product and version you have, and validate any change of product or protocol before patient use. Corrections and additions are welcome — tell us.