Culture dishes, pipettes and plastics — general

Dishes, tubes, pipette tips and stripper tips that contact gametes and embryos.

Clinical review pending·1 min read·Updated 2026-08-16

What it is and how it's used

Any plastic that touches the culture system can be embryotoxic in a bad lot or if not IVF-grade. Changes are easy to overlook because they are 'just consumables'.

Typical failure modes

  • Fertilisation and cleavage drop after a new dish/tip lot or a substitution during a supply gap.
  • Gloves (powdered or new brand) contaminating dishes.

What to check first

  1. Every consumable lot change in the fortnight before the KPI change.
  2. Whether items are MEA/HSSA tested and IVF-labelled.
  3. Glove brand and handling practice.

IFU and practice notes

Use IVF-grade, tested consumables; log lots; avoid substitutions without validation.

Neutral, and the IFU wins

This page summarises common experience across labs and is not manufacturer guidance. Always follow the current instructions for use for the exact product and version you have, and validate any change of product or protocol before patient use. Corrections and additions are welcome — tell us.